Parameter | Condition |
---|---|
Mobile phase 1 (Bottle C) | Acetonitrile |
Mobile phase 2 (Bottle D) | 2 mM ammonium formate |
Mode of mobile phase | Isocratic 50:50 C:D |
Flow rate | 1 mL/min |
Degasser | Normal |
Column temperature | 30°C ± 5 °C |
Autosampler temperature | 20°C ± 5 °C |
Injection volume | 25 μL |
Mode of data acquisition | 3D data collection |
Scan range | 190 to 400 nm |
Resolution | 1.2 |
Sampling rate | 1.0 |
UV max detection for quantification | 210 nm |
Total run timea | 18–25 min |